Title
Construction of a Silver(I) Ion-Selective Electrode Using Amine Phenol Ligand as Carrier and the Selective Determination of Silver in Actual Samples

수송체로서 아민페놀을 이용한 은(I)이온-선택 전극의 제조와 실제 샘플에서 은의 선택적 측정
Author
Wen Ju Xu, Ya Qin Chai, Ruo Yuan*

Wen Ju Xu, Ya Qin Chai, Ruo Yuan*
Keywords
이온-선택 전극, 은(I)이온, 전압 반응, N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane, N,N’-bis(2-hydroxybenzyl)-2,2-dimethyl-1,3- diaminopropane , Ion-selective electrode, Silver(I) ion, Potential response, N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane, N,N’-bis(2-hydr
Abstract
이 작업은 수송체로서 N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane (L1)과 N,N’-bis(2-hydroxybenzyl)-2,2-dimethyl-1,3- diaminopropane (L2)에 기초한 Ag+ 이온-선택 전극의 제조, 개발 그리고 전압반응에 대해 토의한다. Ag+에 대하여 L1에 기 초한 전극은 58.7 mV/dec에 도달하는 안정한 Nernst에 가까운 기울기와 1.0×10-6 M의 검출한계를 갖는 등급의 최소 다섯 가지 순서의 선형범위에서 최적의 전압 반응 특징을 보이는 것을 나타낸다. 제안된 전극은 다른 테스트한 양이온과 비교 하여 Ag+에 우위의 선택성을 보였다. 훌륭한 전압 분석적 특징은 중요한 진짜 샘플에서 은의 순도분석의 성공적인 응용을 이끌 것이다. 그리고 그것은 제안된 Ag+-ISE가 측정 능력의 중요한 진보를 보여준다는 것을 나타내고 있다. 그러나 L2에 기초한 전극에 대해서는, 약한 전압 반응 특징들이 전체 실험 과정에서 관측되었다.

This work discusses the fabrication, development and potential response behaviors of Ag+ ion-selective electrodes (Ag+-ISE) based on N,N’-bis(2-hydroxybenzyl)-1,3-diaminopropane (L1) and N,N’-bis(2-hydroxybenzyl)-2,2-dimethyl-1,3- diaminopropane (L2) as carriers. The observations indicated that the resulting electrode based on L1 toward Ag+ showed stable near-Nernst slope approaching 58.7 mV/dec and the optimum potential response characteristics in a linear range at least five orders of magnitude with a detection limit of 1.0×10-6 M. The proposed electrode displayed the preferential selectivity to Ag+ against other tested cations. The excellent potential analytical characteristics could lead to the successful applications of silver assay in significant real samples, indicating that the proposed Ag+-ISE showed a significant advancement of measurement capabilities. But for the electrode based on L2, the poor potential response characteristics were observed in total experiment process.